Upgrade instructions
To upgrade the Bubbler, connect your PC to the Bubbler using an RS-232 DB9 cable, and use a terminal program.
Microsoft Windows usually comes with a terminal program called HyperTerminal. It can be found by going to the Windows start menu, Programs, Accessories, Communications.
Once started, choose the COM port that you have connected the cable to. Then select these options: 115200 baud, 8 data bits, no parity, 1 stop bit, hardware flow control.
Once connected, press ENTER. Bubbler will respond with a prompt >. You may type VER to see the current software in the sensor.
Type UPG and ENTER to start the download. Bubbler will respond by outputting 'C'. If using HyperTerminal, use the Transfer, Send File menu. Choose the .upg downloaded from this page file and choose Y-Modem.
Once the .upg transfer is complete, hit ENTER again. Type VER to see the new version of the software.
Download CF Bubbler software version 1.23 built in December 2007.
Release Notes - Bubbler - Version 1.23
** Bug
* [BBL-102] - Post purge errors: restrictor autozero range and no pressure increase
* [BBL-123] - Accubar Tank Mismatch mis-reported
* [BBL-124] - Self Test SDI-12 break limits
** Task
* [BBL-120] - Data old issue via SDI-12
* [BBL-126] - Units not reported except for water level
Release Notes - Bubbler - Version 1.22
** Bug
* [BBL-109] - CLONE -XXS command parsing error
* [BBL-110] - Rest not zero error following a leak test
* [BBL-114] - SDI-12 retry required part 2
* [BBL-121] - Display does not turn off on timeout after log download to sd card
** Task
* [BBL-94] - Allow user to name water level measurement
* [BBL-107] - Test Bubbler with Sutron XPert Voice Modem
* [BBL-111] - Change log header
* [BBL-113] - Consolidate averaging times
* [BBL-115] - Add station name to time menu
* [BBL-116] - Tell user to cap outlet when doing leak test via command line
* [BBL-118] - Backlight on whenever the LCD is on
|