Captcha
View our Privacy Policy
Feedback





Captcha
View our Privacy Policy
Feedback

Sea-Bird Coastal Cycle-P Phosphate Sensor

Loading...

Sea-Bird Coastal Cycle-P Phosphate Sensor

    • Robust: Excellent anti-fouling capability provided by copper mesh screens and 10 µm stainless steel intake filters
    • Accurate: Nanomolar resolution and scattering insensitive optical cell
    • Cost Effective: Over 1000 samples including onboard spike calibration for QA/QC, typical deployment duration of three months, reducing field costs

    Mechanical  
    Height (w/handle) 56 cm
    Width 18 cm
    Weight in air 6.8 kg (w/ reagents)
    Weight in water 0.37 kg
    Depth 200 m
    Temperature *1 1 … 35 °C
    Electrical  
    Input  10 … 18 VDC
    Current draw 2.0 A max; 125 mA avg
    Data output  RS-232 or SDI-12
    Host Connector  MCBH-6-MP
    SDI Connector MCBH-8-MP
    Sample rate 2 per hour
    Data memory 1 GB
    Optical  
    LED wavelength 870 nm
    Pathlength 5 cm
    Linearity equal or greater than 95% R2
    Analytical  
    Detection limit, three standard deviations of 18 MOhm water equal to or less than 0.075 µM
    equal to or less than 0.0023 mg/L PO4-P
    Quantification limit, ten standard deviations of 18 MOhm water equal to or less than 0.25 µM
    equal to or less than 0.0077 mg/L PO4-P
    Standard deviation of standard solution of 2.6 µM equal to or less than 0.05 µM
    equal to or less than 0.0015 mg/L PO4-P
    Range, nominal *2 0 … 10 µM
    0-0.3 mg/L PO4-P
    Note: *1 Both storage and operating temperature for sensor and reagent cartridges.
    *2 Higher range is possible (0 … 40 µM, 0 … 1.2 mg/L PO4-P) but is outside specifications
    Units: M = moles per liter.
    mg/L PO4-P refers to mg of atoms phosphorus per liter measured in the form of reactive phosphate.
    ppm ~ mg/L, ppb ~µg/L

    Specifications are subject to change without notice.